ELECTRON MICROSCOPY


Sample

Bacteroides thetaiotaomicron BAM complex

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 105155
Reported Resolution (Å) 3.46
Resolution Method OTHER
Other Details The best parts of two maps at 3.28 A and 3.46 A were combined using phenix.combine_focused_maps
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value 64.9
Fitting Procedure ?
Details Experimental models were fit into the combined focused map.
Data Acquisition
Detector Type FEI FALCON IV (4k x 4k)
Electron Dose (electrons/Å2) 35
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 165000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
MODEL REFINEMENT PHENIX ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?