ELECTRON MICROSCOPY


Sample

70S ribosome

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 196921
Reported Resolution (Å) 2.61
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target ?
Overall B Value 114.14
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON III (4k x 4k)
Electron Dose (electrons/Å2) 28.29
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS GLACIOS
Minimum Defocus (nm) 500
Maximum Defocus (nm) 1500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification 190000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 200
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION cryoSPARC 4.3.1
IMAGE ACQUISITION EPU ?
CTF CORRECTION cryoSPARC 4.4.1
MODEL FITTING Coot 0.9.8.95
INITIAL EULER ASSIGNMENT cryoSPARC 4.3.1
FINAL EULER ASSIGNMENT cryoSPARC 4.4.1
CLASSIFICATION cryoSPARC 4.4.1
RECONSTRUCTION cryoSPARC 4.4.1
MODEL REFINEMENT Servalcat 0.4.77
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?