ELECTRON MICROSCOPY


Sample

Minus end of the vertebrate gamma-tubulin ring complex-capped microtubule

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 8497
Reported Resolution (Å) 23
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space ?
Refinement Protocol ?
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 43
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1000
Maximum Defocus (nm) 3000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 33000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.1
IMAGE ACQUISITION EPU ?
CTF CORRECTION Gctf 1.06
INITIAL EULER ASSIGNMENT RELION 3.1
FINAL EULER ASSIGNMENT RELION 3.1
RECONSTRUCTION RELION 3.1
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?