ELECTRON MICROSCOPY


Sample

Human oxyHb (C1 symmetry) determined using the SPT Labtech chameleon in the presence of 25 uM sodium dithionite under Al's oil

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument SPT LABTECH CHAMELEON
Cryogen Name ETHANE
Sample Vitrification Details Samples were frozen with the SPT Labtech chameleon
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 320799
Reported Resolution (Å) 2.55
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value 89.2
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON IV (4k x 4k)
Electron Dose (electrons/Å2) 60
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS KRIOS
Minimum Defocus (nm) 1000
Maximum Defocus (nm) 2500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 165000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
MODEL REFINEMENT PHENIX 1.21.1_5286
IMAGE ACQUISITION EPU 2
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?