ELECTRON MICROSCOPY


Sample

ZBTB9 BTB domain filament

Specimen Preperation
Sample Aggregation State FILAMENT
Vitrification Instrument LEICA EM GP
Cryogen Name ETHANE
Sample Vitrification Details Grids were vitrified using a Leica EM GP plunge freezer operated at 90% humidity and 10 C with 10 s pre-blot, 3 s blot, 3 s post-blot.
3D Reconstruction
Reconstruction Method HELICAL
Number of Particles 147135
Reported Resolution (Å) 8.1
Resolution Method FSC 0.143 CUT-OFF
Other Details Asymmetric helical refinement followed by local refinement led to the final reconstruction. The helical symmetry values represent the angular rotation and axial rise for two dimers at the core of the map, where the local resolution was the highest. These values, however, were not applied during helical refinement as they did not improve the quality of the final reconstruction.
Refinement Type
Symmetry Type HELICAL
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol OTHER
Refinement Target ?
Overall B Value 762
Fitting Procedure ?
Details Real-space refinement without local grid search
Data Acquisition
Detector Type GATAN K3 (6k x 4k)
Electron Dose (electrons/Å2) 50.1
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TALOS ARCTICA
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 36000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 200
Imaging Details ?
Imaging Experiment
Task Software Package Version
MODEL REFINEMENT PHENIX 1.21.1_5286
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?