ELECTRON MICROSCOPY


Sample

F-actin-Talin(R13-DD) complex

Specimen Preperation
Sample Aggregation State FILAMENT
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details 3.0 ul sample, blotted for 5 s from both sides with filter paper Whatman No.1
3D Reconstruction
Reconstruction Method HELICAL
Number of Particles 206326
Reported Resolution (Å) 2.98
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type HELICAL
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol BACKBONE TRACE
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K3 BIOCONTINUUM (6k x 4k)
Electron Dose (electrons/Å2) 52
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2600
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.1.0
IMAGE ACQUISITION SerialEM ?
CTF CORRECTION CTFFIND 4
MODEL FITTING Coot 0.9.8.6
INITIAL EULER ASSIGNMENT RELION 3.1.0
FINAL EULER ASSIGNMENT RELION 3.1.0
CLASSIFICATION RELION 3.1.0
RECONSTRUCTION RELION 3.1.0
MODEL REFINEMENT PHENIX 1.21-5207
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?