ELECTRON MICROSCOPY


Sample

ACTG2 R257C Filament

Specimen Preperation
Sample Aggregation State FILAMENT
Vitrification Instrument FEI VITROBOT MARK III
Cryogen Name ETHANE
Sample Vitrification Details Blot Force: 0 Blot Time: 2.5 s
3D Reconstruction
Reconstruction Method HELICAL
Number of Particles 473627
Reported Resolution (Å) 2.72
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type HELICAL
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K3 (6k x 4k)
Electron Dose (electrons/Å2) 45
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 500
Maximum Defocus (nm) 2500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 81000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION cryoSPARC 4.0
IMAGE ACQUISITION EPU ?
CTF CORRECTION cryoSPARC 4.0
MODEL FITTING Coot 0.9.8.3
RECONSTRUCTION cryoSPARC 4.0
MODEL REFINEMENT PHENIX 1.20.1
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?