ELECTRON MICROSCOPY


Sample

HIV-1 Nuclear Export complex

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details Washed grid 2x outside of vitrobot; with the second drop of buffer I loaded the tweezers; moved the grid into the chamber and then I blotted.
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 260885
Reported Resolution (Å) 4.25
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space ?
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value 325
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 66
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification 59880
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION ?
PARTICLE SELECTION cryoSPARC ?
CTF CORRECTION cryoSPARC ?
CTF CORRECTION RELION ?
MODEL FITTING UCSF ChimeraX ?
INITIAL EULER ASSIGNMENT RELION ?
INITIAL EULER ASSIGNMENT cryoSPARC ?
FINAL EULER ASSIGNMENT RELION ?
RECONSTRUCTION RELION ?
MODEL REFINEMENT PHENIX ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?