ELECTRON MICROSCOPY


Sample

Ternary Complex of DDB1dB:CRBN:PT-179 with SD40

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument LEICA EM GP
Cryogen Name ETHANE
Sample Vitrification Details Leica EM-GP plunge freezer with chamber conditions of 10 C and 90% relative humidity. Grids were first pre-incubated with 4 uL of 10 uM CRBN-agnostic IKZF1_140-196_Q146A,G151N for 1 minute and then blotted from behind for 4 s. Immediately, 4 uL of mixture 1 diluted 10-fold--with the dilution buffer during the 1-minute incubation time--was applied to the grids before blotting for 4 s and plunging into liquid ethane at -181 C.
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 289691
Reported Resolution (Å) 2.41
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol OTHER
Refinement Target Cross-correlation coefficient
Overall B Value 70
Fitting Procedure ?
Details Phenix real-space refinement without rigid body, with ADP.
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 52
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION Topaz 0.2.5
IMAGE ACQUISITION SerialEM 4.1b
CTF CORRECTION cryoSPARC 4.2.0
MODEL FITTING UCSF ChimeraX 1.6.1
MODEL REFINEMENT PHENIX 1.20.1
INITIAL EULER ASSIGNMENT cryoSPARC 4.2.0
FINAL EULER ASSIGNMENT cryoSPARC 4.2.0
CLASSIFICATION RELION 4.0.1
RECONSTRUCTION cryoSPARC 4.2.0
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?