ELECTRON MICROSCOPY


Sample

Outer kinetochore Ndc80-Nuf2 CH domains bound to Dam1 in complex with alpha-beta tubulin

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument ?
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 95730
Reported Resolution (Å) 3.54
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol OTHER
Refinement Target ?
Overall B Value 540.98
Fitting Procedure ?
Details Initial rigid body fitting was performed in chimera, with manual correction in coot and real-space refinement in PHENIX
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 50
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS KRIOS
Minimum Defocus (nm) 1200
Maximum Defocus (nm) 3000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification ?
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
MODEL REFINEMENT PHENIX 1.20.1_4487:
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?