ELECTRON MICROSCOPY


Sample

RBD_Fab_AM032_AM047 complex

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details 2.5 ul of samples placed on grid before plunge frozen with 5s blot time
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 387000
Reported Resolution (Å) 3.3
Resolution Method FSC 0.143 CUT-OFF
Other Details Final map generated from 3D flex reconstruction following non-uniform refinement.
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 50
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.73
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification 105000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION cryoSPARC 3.3.2
IMAGE ACQUISITION EPU ?
CTF CORRECTION cryoSPARC 3.3.2
MODEL FITTING UCSF Chimera 1.16
MODEL REFINEMENT PHENIX 1.21rc1_4895:
MODEL REFINEMENT Coot 0.9.6 EL
INITIAL EULER ASSIGNMENT cryoSPARC 3.3.2
FINAL EULER ASSIGNMENT cryoSPARC 3.3.2
CLASSIFICATION cryoSPARC 4.0.0
RECONSTRUCTION cryoSPARC 4.1.0 and 4.2.1
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?