ELECTRON MICROSCOPY


Sample

Folate-specific ECF transporter complex

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE-PROPANE
Sample Vitrification Details 5 mM Mg-AMP-PNP and 10 uM folate were added to the concentrated sample and incubated at 37 degC for 10 min. 2.9 mM fluorinated Fos-choline 8 was added prior to sample application onto grids. Grids were blotted for 3.5-4 sec.
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 129663
Reported Resolution (Å) 3.6
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 50.1
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TALOS ARCTICA
Minimum Defocus (nm) 800
Maximum Defocus (nm) 1800
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 130000
Calibrated Magnification 48924
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 200
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION crYOLO 1.7.5
IMAGE ACQUISITION SerialEM 3.8.0 beta
CTF CORRECTION CTFFIND 4.1.14
MODEL FITTING Coot 0.9.8.1
MODEL REFINEMENT PHENIX 1.20.1-4487-000
INITIAL EULER ASSIGNMENT cryoSPARC 3.3.2
FINAL EULER ASSIGNMENT cryoSPARC 3.3.2
RECONSTRUCTION cryoSPARC 3.3.2
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?