ELECTRON MICROSCOPY


Sample

Dicing state of mouse oocyte dicer with pre-miR-15a

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details Described in STAR methods
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 84697
Reported Resolution (Å) 6.21
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target Correlation coefficient
Overall B Value 451.4
Fitting Procedure ?
Details Initial model for mouse oocyte Dicer was calculated using AlphaFold 2.0. Initial rigid body fitting was done using Chimera. Further refinement was done using ISOLDE with distance restrains
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 55
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1000
Maximum Defocus (nm) 3500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification ?
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION crYOLO 1.7.5
CTF CORRECTION cryoSPARC v3.3.1
MODEL FITTING UCSF ChimeraX 1.15
CLASSIFICATION cryoSPARC v3.3.1
RECONSTRUCTION cryoSPARC v3.3.1
MODEL REFINEMENT ISOLDE 1.1.0
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?