ELECTRON MICROSCOPY


Sample

Lloviu cuevavirus nucleoprotein RNA complex

Specimen Preperation
Sample Aggregation State HELICAL ARRAY
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details Apply 1.25 micro litters of sample from each side of the grid and blot for 7 seconds before plugging
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 26797
Reported Resolution (Å) 3.0356
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol AB INITIO MODEL
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON III (4k x 4k)
Electron Dose (electrons/Å2) 40
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 1800
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 0.01
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 59000
Calibrated Magnification 59000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.0
IMAGE ACQUISITION EPU ?
CTF CORRECTION CTFFIND 4.1.14
MODEL FITTING Coot 0.8.9.2
INITIAL EULER ASSIGNMENT RELION 3.0
FINAL EULER ASSIGNMENT RELION 3.0
CLASSIFICATION RELION 3.0
RECONSTRUCTION RELION 3.0
MODEL REFINEMENT PHENIX 1.18.2
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?