ELECTRON MICROSCOPY


Sample

V1EG domain of V/A-ATPase from Thermus thermophilus at low ATP concentration, state1-2

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 35827
Reported Resolution (Å) 3.0
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol OTHER
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details The atomic model built in this study was used as an initial model.
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 50
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 800
Maximum Defocus (nm) 2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 0.071
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 81000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.1
IMAGE ACQUISITION SerialEM ?
CTF CORRECTION CTFFIND 4.1
MODEL FITTING Coot ?
INITIAL EULER ASSIGNMENT cryoSPARC 3.2
FINAL EULER ASSIGNMENT RELION 3.1
CLASSIFICATION RELION 3.1
RECONSTRUCTION RELION 3.1
MODEL REFINEMENT PHENIX ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?