ELECTRON MICROSCOPY


Sample

Structure of the Caulobacter crescentus S-layer protein RsaA N-terminal domain bound to LPS and soaked with Holmium

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details Vitrobot options: Blot time 4 seconds, Blot force -13,1, Wait time 10 seconds, Drain time 0.5 seconds,
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 158430
Reported Resolution (Å) 4.37
Resolution Method FSC 0.143 CUT-OFF
Other Details The final map was obtained from 158,430 particles and post-processed using a soft mask focused on the inner fourteen subunits yielding a resolution of 4.37 A according to the gold standard Fourier shell correlation criterion of 0.143 (Scheres, 2012) with some anisotropy in Z as judged by directional FSCs (Tan et al., 2017)
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details The atomic coordinates (PDB ID 6T72) of our previous cryo-EM structure (von Kugelgen et al., 2020) of the RsaANTD oligomer bound to the O-antigen of lipopolysaccharide (LPS) were rigid body fitted into the final post-processed map from Relion 3.0 (Zivanov et al., 2018) using UCSF Chimera (Pettersen et al., 2004). The resulting fitted model was subjected to real-space refinement using Refmac5 (Murshudov et al., 2011) inside the CCP-EM suite (Burnely et al., 2017), as described previously (von Kugelgen et al., 2020), using reference restraints of the initial structure (PDB ID 6T72) generated with PROSMART (Nicholls et al. 2012).
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 44.8
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) -1000
Maximum Defocus (nm) -4000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 130000
Calibrated Magnification 130000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details EPU software
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.0
IMAGE ACQUISITION EPU ?
CTF CORRECTION CTFFIND 4.1.13
MODEL FITTING UCSF Chimera 1.13
INITIAL EULER ASSIGNMENT RELION 3.0
FINAL EULER ASSIGNMENT RELION 3.0
CLASSIFICATION RELION 3.0
RECONSTRUCTION RELION 3.0
MODEL REFINEMENT REFMAC 5.8.0258
MODEL REFINEMENT PHENIX ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION RELION refinement with in-built CTF correction. The function is similar to a Wiener filter, so amplitude correction included.