7N5D

ELECTRON MICROSCOPY


Sample

FLYC1

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 797000
Reported Resolution (Å) 2.8
Resolution Method OTHER
Other Details Composite model generated in Chimera
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol AB INITIO MODEL
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 50
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 700
Maximum Defocus (nm) 2200
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 29000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION Leginon ?
CTF CORRECTION Gctf 1.06
CTF CORRECTION CTFFIND 4
MODEL FITTING Coot 0.9
INITIAL EULER ASSIGNMENT cryoSPARC 2
FINAL EULER ASSIGNMENT RELION 3.1
CLASSIFICATION RELION 3.1
RECONSTRUCTION UCSF Chimera 1.14
MODEL REFINEMENT Coot 0.9
MODEL REFINEMENT PHENIX 1.18.2-3874
MODEL REFINEMENT Rosetta 3.10
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?