ELECTRON MICROSCOPY


Sample

infectious mammalian prion fibril (263K scrapie)

Specimen Preperation
Sample Aggregation State FILAMENT
Vitrification Instrument LEICA EM GP
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 15884
Reported Resolution (Å) 3.14
Resolution Method FSC 0.143 CUT-OFF
Other Details Auto-refinement was then performed while optimizing the helical twist and rise. Auto-refinement with refinement of twist and rise yielded a final map with a twist of -0.847 degree and rise of 4.874 angstrom. Iterative cycles of CTF refinement, Bayesian polishing, and auto refinement were used until resolution estimates stabilized. Post processing in RELION was performed with a soft-edged mask representing 10% of the central Z length of the fibril. Sharpening was applied with a B-factor of -31 square angstrom.
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space RECIPROCAL
Refinement Protocol AB INITIO MODEL
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 60
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification ?
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.1
IMAGE ACQUISITION SerialEM ?
CTF CORRECTION CTFFIND 4.1
MODEL FITTING Coot 0.9.2pre
MODEL FITTING PHENIX 1.19
MODEL REFINEMENT PHENIX 1.19
MODEL REFINEMENT REFMAC 5
INITIAL EULER ASSIGNMENT RELION 3.1
FINAL EULER ASSIGNMENT RELION 3.1
CLASSIFICATION RELION 3.1
RECONSTRUCTION RELION 3.1
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?