ELECTRON MICROSCOPY


Sample

porcine native cardiac thin filament decorated with porcine cardiac myosin-S1

Specimen Preperation
Sample Aggregation State HELICAL ARRAY
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method HELICAL
Number of Particles 115746
Reported Resolution (Å) 3.8
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type HELICAL
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details SWISS-MODEL web service used to create initial homology models of cardiac actin and myosin
Data Acquisition
Detector Type FEI FALCON III (4k x 4k)
Electron Dose (electrons/Å2) 20
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification ?
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION EMAN ?
CTF CORRECTION CTFFIND 4
MODEL FITTING DireX ?
MODEL REFINEMENT PHENIX 1.17
INITIAL EULER ASSIGNMENT RELION 3.08
FINAL EULER ASSIGNMENT RELION 3.08
CLASSIFICATION RELION 3.08
RECONSTRUCTION RELION 3.08
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?