ELECTRON MICROSCOPY


Sample

Mature 50S ribosomal subunit from RrmJ knock out E.coli strain

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details blot for 2 seconds before plunging
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 98194
Reported Resolution (Å) 3.14
Resolution Method FSC 0.143 CUT-OFF
Other Details A soft mask was added to improve resolution at the final refinement.
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space RECIPROCAL
Refinement Protocol RIGID BODY FIT
Refinement Target Correlation coefficient
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON II (4k x 4k)
Electron Dose (electrons/Å2) 45
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) -1000
Maximum Defocus (nm) -2000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 75000
Calibrated Magnification 75000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details Preliminary grid screening was performed manually.
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 1.4
CTF CORRECTION CTFFIND 3.0
MODEL FITTING UCSF Chimera 1.13
INITIAL EULER ASSIGNMENT RELION 1.4
FINAL EULER ASSIGNMENT RELION 1.4
CLASSIFICATION RELION 1.4
RECONSTRUCTION RELION 1.4
MODEL REFINEMENT PHENIX 1.16
MODEL REFINEMENT REFMAC 5
MODEL FITTING UCSF Chimera 1.13
MODEL REFINEMENT PHENIX 1.16
MODEL REFINEMENT REFMAC 5
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION CTF amplitude correction was performed following 3D reconstruction.