ELECTRON MICROSCOPY


Sample

Complement C4b bound to nanobody E3

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details blot 4.5 seconds, force 1
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 176318
Reported Resolution (Å) 3.39
Resolution Method FSC 0.143 CUT-OFF
Other Details performed in RELION
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol ?
Refinement Target Map to model FSC
Overall B Value ?
Fitting Procedure ?
Details Model was refined in the local-resolution filtered Relion map using Phenix real space refine.
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 51.7
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TALOS ARCTICA
Minimum Defocus (nm) 800
Maximum Defocus (nm) 3000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 130000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 200
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 3.1
IMAGE ACQUISITION EPU ?
CTF CORRECTION Gctf 1.18
MODEL FITTING UCSF Chimera 1.12
INITIAL EULER ASSIGNMENT RELION 3.1
FINAL EULER ASSIGNMENT RELION 3.1
CLASSIFICATION RELION 3.1
RECONSTRUCTION RELION 3.1
MODEL REFINEMENT PHENIX 1.18.2-3874
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?