ELECTRON MICROSCOPY


Sample

Pseudomonas virus Pf1

Specimen Preperation
Sample Aggregation State FILAMENT
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details Samples for cryo-EM were prepared by pipetting 2.5 ul of the sample onto freshly glow-discharged Quantifoil grids (Cu/Rh R2/2, 200 mesh). Grids were blotted for 2.5 seconds with a blot force of -15, 0.5 second drain and 0 second wait times.
3D Reconstruction
Reconstruction Method HELICAL
Number of Particles 95481
Reported Resolution (Å) 3.9
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type HELICAL
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol BACKBONE TRACE
Refinement Target Correlation coefficient
Overall B Value 180.97
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 43
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) -1000
Maximum Defocus (nm) -3000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification 105000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 2.0
IMAGE ACQUISITION EPU ?
CTF CORRECTION RELION 2.0
MODEL FITTING Coot ?
INITIAL EULER ASSIGNMENT RELION 2.0
FINAL EULER ASSIGNMENT RELION 2.0
CLASSIFICATION RELION 3.0
RECONSTRUCTION RELION 3.0
MODEL REFINEMENT PHENIX ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION Wiener filter implemented in the RELION refinement algorithm