ELECTRON MICROSCOPY


Sample

Body domain of the mt-SSU assemblosome

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE-PROPANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 161661
Reported Resolution (Å) 3.1
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space ?
Refinement Protocol OTHER
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details Starting models of ribosomal proteins of the T. brucei mitochondrial ribosomal small subunit were taken from PDB 6HIW. Mitchondrial ribosomal small subunit assembly factors were built de novo.
Data Acquisition
Detector Type FEI FALCON III (4k x 4k)
Electron Dose (electrons/Å2) 40
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 59000
Calibrated Magnification 100719
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION EPU ?
CTF CORRECTION Gctf ?
INITIAL EULER ASSIGNMENT RELION ?
FINAL EULER ASSIGNMENT RELION ?
RECONSTRUCTION RELION ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION On the fly in RELION