ELECTRON MICROSCOPY


Sample

Cytochrome bd-I oxidase from E. coli

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 170000
Reported Resolution (Å) 2.68
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol AB INITIO MODEL
Refinement Target Cross-correlation coefficient
Overall B Value 120
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON III (4k x 4k)
Electron Dose (electrons/Å2) 40
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 500
Maximum Defocus (nm) 2500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 96000
Calibrated Magnification 96000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION EPU ?
CTF CORRECTION CTFFIND 4
MODEL FITTING Coot 0.8
INITIAL EULER ASSIGNMENT RELION 3
FINAL EULER ASSIGNMENT RELION 3
CLASSIFICATION RELION 3
RECONSTRUCTION RELION 3
MODEL REFINEMENT PHENIX 1.14
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
NONE ?