ELECTRON MICROSCOPY


Sample

ExeD

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 51960
Reported Resolution (Å) 3.7
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol OTHER
Refinement Target Cross-correlation coefficient
Overall B Value 182
Fitting Procedure ?
Details Coot and buccaneer for model building and refmac5 for refinement (CCPEM-1.1.0 suite).
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 40
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI POLARA 300
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.0
Imaging Mode BRIGHT FIELD
Specimen Holder Model GATAN LIQUID NITROGEN
Nominal Magnification ?
Calibrated Magnification 41322
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION Latitude S
CTF CORRECTION Gctf ?
CTF CORRECTION RELION 2.1
MODEL FITTING Coot ?
INITIAL EULER ASSIGNMENT RELION 2.1
FINAL EULER ASSIGNMENT RELION 2.1
CLASSIFICATION RELION 2.1
RECONSTRUCTION RELION 2.1
MODEL REFINEMENT REFMAC ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?