ELECTRON MICROSCOPY


Sample

PTC3 holotoxin complex in prepore state - Mutant TcC-D651A

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument GATAN CRYOPLUNGE 3
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 132000
Reported Resolution (Å) 3.72
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target Model to Map FSC
Overall B Value 49.43
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON II (4k x 4k)
Electron Dose (electrons/Å2) 2.2
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 0
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 59000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION EMAN2 ?
IMAGE ACQUISITION EPU ?
CTF CORRECTION SPHIRE ?
MODEL FITTING UCSF Chimera 1.12
INITIAL EULER ASSIGNMENT SPHIRE ?
FINAL EULER ASSIGNMENT SPHIRE ?
CLASSIFICATION SPHIRE ?
RECONSTRUCTION SPHIRE ?
MODEL REFINEMENT Rosetta ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?