ELECTRON MICROSCOPY


Sample

Tau filaments extracted from the frontotemporal cortex of a patient with Pick's disease

Specimen Preperation
Sample Aggregation State TISSUE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method HELICAL
Number of Particles 16097
Reported Resolution (Å) 3.2
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type HELICAL
Map-Model Fitting and Refinement
ID 1
Refinement Space RECIPROCAL
Refinement Protocol AB INITIO MODEL
Refinement Target Fourier shell correlation
Overall B Value 57
Fitting Procedure ?
Details Fourier-space refinement of the complete atomic model against the narrow Pick filament map was performed in REFMAC. A stack of three consecutive monomers was refined to preserve nearest-neighbour interactions for the middle chain.
Data Acquisition
Detector Type GATAN K2 QUANTUM (4k x 4k)
Electron Dose (electrons/Å2) 1.06
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1700
Maximum Defocus (nm) 2800
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification ?
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION EPU ?
CTF CORRECTION Gctf 1.06
MODEL FITTING Coot 0.8.9.1
RECONSTRUCTION RELION 2.1
MODEL REFINEMENT REFMAC 5.7.0032
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?