6G2J

ELECTRON MICROSCOPY


Sample

Mitochondrial complex I from mouse in the active state

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details 8-16s blotting.
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 20370
Reported Resolution (Å) 3.3
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details Used default paramteres in phenix.real_space_refine
Data Acquisition
Detector Type GATAN K2 QUANTUM (4k x 4k)
Electron Dose (electrons/Å2) 50.0
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1500
Maximum Defocus (nm) 3100
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification ?
Calibrated Magnification 47600
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION EPU 1.9
CTF CORRECTION Gctf 1.06
MODEL FITTING Coot 0.8.9.1
MODEL REFINEMENT PHENIX 1.13
INITIAL EULER ASSIGNMENT RELION 2.1
FINAL EULER ASSIGNMENT RELION 2.1
CLASSIFICATION RELION 2.1
RECONSTRUCTION RELION 2.1
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?