ELECTRON MICROSCOPY


Sample

Target of rapamycin protein complex 2

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details 2 - 3 sec blotting
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 16190
Reported Resolution (Å) 7.9
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type FEI FALCON II (4k x 4k)
Electron Dose (electrons/Å2) 50
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1500
Maximum Defocus (nm) 3500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION SerialEM 3.6
CTF CORRECTION CTFFIND 4
MODEL FITTING Coot 0.8.8
INITIAL EULER ASSIGNMENT RELION 1.4
FINAL EULER ASSIGNMENT RELION 1.4
RECONSTRUCTION RELION 1.4
PARTICLE SELECTION EMAN2 ?
CTF CORRECTION CTFFIND 4
INITIAL EULER ASSIGNMENT RELION 1.4
FINAL EULER ASSIGNMENT RELION 1.4
RECONSTRUCTION RELION 1.4
MODEL REFINEMENT REFMAC 5.8.0158
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details