ELECTRON MICROSCOPY


Sample

Dicer and trbp

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 130000
Reported Resolution (Å) 4.7
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target CC
Overall B Value 100
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 50.0
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 2500
Maximum Defocus (nm) 3500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 0.01
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 89000
Calibrated Magnification 36496
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details no
Imaging Experiment
Task Software Package Version
CTF CORRECTION RELION 2.1
MODEL FITTING UCSF Chimera 1.10
MODEL REFINEMENT PHENIX 1.11
INITIAL EULER ASSIGNMENT RELION 2.1
FINAL EULER ASSIGNMENT RELION 2.1
CLASSIFICATION RELION 2.1
RECONSTRUCTION RELION 2.1
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION ?