ELECTRON MICROSCOPY


Sample

Complex of zebrafish (D. rerio) STRA6 with copurified calmodulin reconstituted in amphipol A8-35

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details 3 uL sample applied, 3-4 second blot time, 30 second wait time, blot force 3, grid blotted from both sides, plunged into liquid ethane (FEI VITROBOT MARK IV).
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 56615
Reported Resolution (Å) 3.9
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space ?
Refinement Protocol AB INITIO MODEL
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 100
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TECNAI F30
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification ?
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION RELION 1.3
IMAGE ACQUISITION Leginon ?
CTF CORRECTION CTFFIND4 4.0.17
MODEL FITTING Coot 0.8.4
FINAL EULER ASSIGNMENT RELION 1.3
RECONSTRUCTION RELION 1.3
MODEL REFINEMENT PHENIX dev-2481