5I68

ELECTRON MICROSCOPY


Sample

alcohol oxidase

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK I
Cryogen Name ETHANE
Sample Vitrification Details blot for 11 seconds before plunging
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 49559
Reported Resolution (Å) 3.37
Resolution Method FSC 0.143 CUT-OFF
Other Details RELION was used for the reconstruction
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol AB INITIO MODEL
Refinement Target ?
Overall B Value 147
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 51
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model JEOL 3200FSC
Minimum Defocus (nm) ?
Maximum Defocus (nm) ?
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 4.2
Imaging Mode BRIGHT FIELD
Specimen Holder Model JEOL 3200FSC CRYOHOLDER
Nominal Magnification 30000
Calibrated Magnification 43860
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details Data was collected manually
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION EMAN ?
CTF CORRECTION RELION 1.3
MODEL FITTING UCSF Chimera ?
INITIAL EULER ASSIGNMENT RELION 1.3
FINAL EULER ASSIGNMENT RELION 1.3
CLASSIFICATION RELION 1.3
RECONSTRUCTION RELION 1.3
MODEL REFINEMENT Coot ?
MODEL REFINEMENT PHENIX ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING ONLY CTF was determined by CTFFIND3 inside the RELION software