X-RAY DIFFRACTION

Crystallization Details
Method pH Temprature Details
X-RAY DIFFRACTION 7.8 298 25% PEG 8K, 70 mM NaCl, 100 mM Tris-HCl pH 7.8
Unit Cell:
a: 68.358 Å b: 159.696 Å c: 197.360 Å α: 90.00° β: 90.00° γ: 90.00°
Symmetry:
Space Group: P 21 21 21
Crystal Properties:
Matthew's Coefficient: 2.55 Solvent Content: 51.69
Refinement Statistics
Diffraction ID Structure Solution Method Cross Validation Method Resolution Limit (High) Resolution Limit (Low) Number of Reflections (Observed) Number of Reflections (R-free) Percent Reflections (Observed) R-Work R-Free Mean Isotropic
X-RAY DIFFRACTION MOLECULAR REPLACEMENT FREE R-VALUE 2.105 41.973 124273 6242 98.71 0.1986 0.2364 ?
Data Collection
Overall
Resolution Limit (High) Resolution Limit (Low) Percent Possible (Observed) R Merge I (Observed) R Sym I (Observed) Net I Over Average Sigma (I) Redundancy Number Reflections (All) Number Reflections (Observed) Observed Criterion Sigma (F) Observed Criterion Sigma (I) B (Isotropic) From Wilson Plot
2.1 50 99.2 0.156 ? 10.1 4.7 ? 125484 ? ? ?
Highest Resolution Shell
# Resolution Limits (Low) Resolution Limits (High) Percent Possible (All) Percent Possible (Observed) R-Sym I (Observed) Mean I Over Sigma (Observed) Redundancy Number Unique Reflections (All)
1 2.1 2.18 97.8 ? ? 1.74 4.7 ?
Diffraction
Diffraction experiment
Crystal ID Scattering Type Data Collection Temprature Detector Detector Type Details Collection Date Monochromator Protocol
1 100 K ?
Radiation Source
Source Type Wavelength List Synchrotron Site Beamline
SYNCHROTRON SSRF BEAMLINE BL17U 0.98 SSRF BL17U
Software
Software Name Purpose Version
PHENIX refinement 1.9_1692
HKL-2000 data reduction .
HKL-2000 data scaling .
PHENIX phasing .
Feedback Form
Name
Email
Institute
Feedback