ELECTRON MICROSCOPY


Sample

BOVINE MITOCHONDRIAL ATP SYNTHASE

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK III
Cryogen Name ETHANE-PROPANE
Sample Vitrification Details VITRIFICATION 1 -- CRYOGEN- ETHANE-PROPANE MIXTURE, HUMIDITY- 100, INSTRUMENT- FEI VITROBOT MARK III, METHOD- BLOT FOR 27 SECONDS BEFORE PLUNGING,
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 22117
Reported Resolution (Å) 7.1
Resolution Method ?
Other Details SUBMISSION BASED ON EXPERIMENTAL DATA FROM EMDB EMD-3170. (DEPOSITION ID: 13802).
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space ?
Refinement Protocol FLEXIBLE FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details METHOD--FLEXIBLE FITTING
Data Acquisition
Detector Type GATAN K2 (4k x 4k)
Electron Dose (electrons/Å2) 60.3
Imaging Experiment
Date of Experiment 2015-03-15
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1200
Maximum Defocus (nm) 4100
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model .
Nominal Magnification 18000
Calibrated Magnification 30487
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details K2 SUMMIT DIRECT DETECTOR DEVICE (GATAN INC.) OPERATED IN SUPER-RESOLUTION MODE WITH A 1.64 ANGSTROM PHYSICAL PIXEL AND 0. 82 ANGSTROM SUPER- RESOLUTION PIXEL. WITH NO SPECIMEN PRESENT, THE RATE OF EXPOSURE OF THE DETECTOR WAS 8 ELECTRONS PER PIXEL PER SECOND. EXPOSURE- FRACTIONATED MOVIES OF 20.1 S WERE RECORDED AS STACKS OF 67 FRAMES, SO THAT SELECTED SPECIMEN AREAS WERE EXPOSED WITH A TOTAL OF 60.3 ELECTRONS PER SQUARE ANGSTROM.
Imaging Experiment
Task Software Package Version
MODEL FITTING MDFF ?
MODEL FITTING UCSF Chimera ?
RECONSTRUCTION FREALIGN ?
RECONSTRUCTION RELION ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
? EACH PARTICLE