X-RAY DIFFRACTION

Crystallization Details
Method pH Temprature Details
X-RAY DIFFRACTION 7.0 293 31% PEG 6000, 100MM MGCL2, 8MM BETA MERCEPTOETHANOL, 100MM TRIS HCL
Unit Cell:
a: 36.907 Å b: 58.407 Å c: 63.808 Å α: 90.000° β: 90.000° γ: 90.000°
Symmetry:
Space Group: P 21 21 21
Crystal Properties:
Matthew's Coefficient: 2.20 Solvent Content: 44.19
Refinement Statistics
Diffraction ID Structure Solution Method Cross Validation Method Resolution Limit (High) Resolution Limit (Low) Number of Reflections (Observed) Number of Reflections (R-free) Percent Reflections (Observed) R-Work R-Free Mean Isotropic
X-RAY DIFFRACTION MOLECULAR REPLACEMENT THROUGHOUT 1.5000 43.0800 20313 1103 94.5200 0.1593 0.1753 18.0760
Data Collection
Overall
Resolution Limit (High) Resolution Limit (Low) Percent Possible (Observed) R Merge I (Observed) R Sym I (Observed) Net I Over Average Sigma (I) Redundancy Number Reflections (All) Number Reflections (Observed) Observed Criterion Sigma (F) Observed Criterion Sigma (I) B (Isotropic) From Wilson Plot
1.50 43.083 94.900 ? 0.058 14.900 5.200 21521 21521 ? ? ?
Highest Resolution Shell
# Resolution Limits (Low) Resolution Limits (High) Percent Possible (All) Percent Possible (Observed) R-Sym I (Observed) Mean I Over Sigma (Observed) Redundancy Number Unique Reflections (All)
1 1.500 1.580 67.400 ? 0.251 3.000 1.600 2152
Diffraction
Diffraction experiment
Crystal ID Scattering Type Data Collection Temprature Detector Detector Type Details Collection Date Monochromator Protocol
1 293 K ?
Radiation Source
Source Type Wavelength List Synchrotron Site Beamline
ROTATING ANODE BRUKER AXS MICROSTAR 1.54184 ? ?
Software
Software Name Purpose Version
SAINT data reduction .
SCALA data scaling 3.3.16
REFMAC refinement 5.5.0109
PDB_EXTRACT data extraction 3.15
REFMAC phasing 5.5.0109