ELECTRON MICROSCOPY


Sample

HCV-IRES 80S INITIATION COMPLEX WITH EIF5B POST- LIKE STATE

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK I
Cryogen Name ETHANE
Sample Vitrification Details VITRIFICATION 1 -- CRYOGEN- ETHANE, HUMIDITY- 100, TEMPERATURE- 93, INSTRUMENT- FEI VITROBOT MARK I, METHOD- BLOT FOR 2-4 SECONDS BEFORE PLUNGING,
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 541570
Reported Resolution (Å) 9.5
Resolution Method ?
Other Details SUBMISSION BASED ON EXPERIMENTAL DATA FROM EMDB EMD-2683. (DEPOSITION ID: 12564).
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details METHOD--RIGID BODY
Data Acquisition
Detector Type TVIPS TEMCAM-F416 (4k x 4k)
Electron Dose (electrons/Å2) 20
Imaging Experiment
Date of Experiment 2012-12-22
Temprature (Kelvin)
Microscope Model FEI TECNAI F30
Minimum Defocus (nm) 2000
Maximum Defocus (nm) 4500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.0
Imaging Mode BRIGHT FIELD
Specimen Holder Model .
Nominal Magnification 115000
Calibrated Magnification 194805
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details AUTOMATED DATA COLLECTION USING LEGINON
Imaging Experiment
Task Software Package Version
RECONSTRUCTION SPARX ?
RECONSTRUCTION SPIDER ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
? CTFFIND3