ELECTRON MICROSCOPY


Sample

DENGUE VIRUS SEROTYPE 1 WITH FAB FRAGMENTS OF HUMAN MONOCLONAL ANTIBODY 1F4

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK IV
Cryogen Name ETHANE
Sample Vitrification Details VITRIFICATION 1 -- CRYOGEN- ETHANE, HUMIDITY- 100, TEMPERATURE- 100, INSTRUMENT- FEI VITROBOT MARK IV, METHOD- BLOT WITH FILTER PAPER FOR 2 S BEFORE PLUNGING,
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 10270
Reported Resolution (Å) 6
Resolution Method ?
Other Details SUBMISSION BASED ON EXPERIMENTAL DATA FROM EMDB EMD-2442. (DEPOSITION ID: 11902).
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target Cross-correlation coefficient
Overall B Value ?
Fitting Procedure ?
Details METHOD--RIGID BODY REFINEMENT PROTOCOL--CRYO-EM
Data Acquisition
Detector Type FEI FALCON I (4k x 4k)
Electron Dose (electrons/Å2) 17.5
Imaging Experiment
Date of Experiment 2012-09-14
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1000
Maximum Defocus (nm) 4000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model .
Nominal Magnification 47000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
RECONSTRUCTION EMAN ?
RECONSTRUCTION MPSA ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
? EACH PARTICLE