ELECTRON MICROSCOPY


Sample

MS2 BOUND TO F-PILUS

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument HOMEMADE PLUNGER
Cryogen Name ETHANE
Sample Vitrification Details LIQUID ETHANE HOME MADE FREEZING DEVICE
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 1500
Reported Resolution (Å) 39
Resolution Method ?
Other Details THREE COORDINATE FILES (4BP4, 4BP5, 4BP6) LINKED TO 1 EM STRUCTURE EMD-2365.
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details METHOD--RIGID BODY REFINEMENT PROTOCOL--RIGID BODY
Data Acquisition
Detector Type GATAN ULTRASCAN 10000 (10k x 10k)
Electron Dose (electrons/Å2) ?
Imaging Experiment
Date of Experiment 2012-01-01
Temprature (Kelvin)
Microscope Model FEI TECNAI 12
Minimum Defocus (nm) 3000
Maximum Defocus (nm) 3000
Minimum Tilt Angle (degrees) -60
Maximum Tilt Angle (degrees) 60
Nominal CS ?
Imaging Mode BRIGHT FIELD
Specimen Holder Model .
Nominal Magnification 23000
Calibrated Magnification 23000
Source LAB6
Acceleration Voltage (kV) 120
Imaging Details ?
Imaging Experiment
Task Software Package Version
RECONSTRUCTION IMOD ?
RECONSTRUCTION PEET ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details