X-RAY DIFFRACTION

Crystallization Details
Method pH Temprature Details
X-RAY DIFFRACTION 8.1 295.5 1.25 Sodium Citrate, 0.1M HEPES, pH 8.1, vapor diffusion, sitting drop, temperature 295.5K
Unit Cell:
a: 38.629 Å b: 62.318 Å c: 66.034 Å α: 64.250° β: 80.040° γ: 72.120°
Symmetry:
Space Group: P 1
Crystal Properties:
Matthew's Coefficient: 2.77 Solvent Content: 55.60
Refinement Statistics
Diffraction ID Structure Solution Method Cross Validation Method Resolution Limit (High) Resolution Limit (Low) Number of Reflections (Observed) Number of Reflections (R-free) Percent Reflections (Observed) R-Work R-Free Mean Isotropic
X-RAY DIFFRACTION MOLECULAR REPLACEMENT THROUGHOUT 1.6500 36.7200 60214 3056 ? 0.2010 0.2335 26.9027
Data Collection
Overall
Resolution Limit (High) Resolution Limit (Low) Percent Possible (Observed) R Merge I (Observed) R Sym I (Observed) Net I Over Average Sigma (I) Redundancy Number Reflections (All) Number Reflections (Observed) Observed Criterion Sigma (F) Observed Criterion Sigma (I) B (Isotropic) From Wilson Plot
1.650 50.000 94.800 0.053 ? 12.000 1.900 ? 60290 ? ? 22.6
Highest Resolution Shell
# Resolution Limits (Low) Resolution Limits (High) Percent Possible (All) Percent Possible (Observed) R-Sym I (Observed) Mean I Over Sigma (Observed) Redundancy Number Unique Reflections (All)
1 1.650 1.710 83.800 ? ? ? 1.600 5312
Diffraction
Diffraction experiment
Crystal ID Scattering Type Data Collection Temprature Detector Detector Type Details Collection Date Monochromator Protocol
1 100 K
Radiation Source
Source Type Wavelength List Synchrotron Site Beamline
SYNCHROTRON APS BEAMLINE 23-ID-B 0.97958 APS 23-ID-B
Software
Software Name Purpose Version
DENZO data reduction .
SCALEPACK data scaling .
PHASER phasing 1.3.3
BUSTER-TNT refinement .
PDB_EXTRACT data extraction 3.10
HKL-2000 data collection .
HKL-2000 data reduction .
HKL-2000 data scaling .
BUSTER refinement 2.8.0
Feedback Form