ELECTRON MICROSCOPY


Sample

Rat TRPV1

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument FEI VITROBOT MARK III
Cryogen Name ETHANE
Sample Vitrification Details Blot for 6 seconds before plunging into liquid ethane (FEI VITROBOT MARK III)
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 35645
Reported Resolution (Å) 3.275
Resolution Method FSC 0.143 CUT-OFF
Other Details The entire ankyrin structure was fitted as a single rigid body. As a result, the first two ankyrin repeats (UNP residues 111-199) are included but are not well defined by the experimental density. (Single particle details: 3D classification, refinement, and reconstruction were performed using RELION) (Single particle--Applied symmetry: C4)
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol AB INITIO MODEL
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details REFINEMENT PROTOCOL--de novo model building DETAILS--The entire ankyrin structure was fitted as a single rigid body. As a result, the first two ankyrin repeats (UNP residues 111-199) are included but are not well defined by the experimental density.
Data Acquisition
Detector Type GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å2) 21
Imaging Experiment
Date of Experiment 2013-01-01
Temprature (Kelvin)
Microscope Model FEI POLARA 300
Minimum Defocus (nm) 1500
Maximum Defocus (nm) 3000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.0
Imaging Mode BRIGHT FIELD
Specimen Holder Model OTHER
Nominal Magnification 31000
Calibrated Magnification 31000
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details K2 Summit in super-resolution counting mode. Motion correction as described in Li et al. (2013) Nature Methods.
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
. Each particle