ELECTRON MICROSCOPY


Sample

Sputnik virus

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument GATAN CRYOPLUNGE 3
Cryogen Name ETHANE
Sample Vitrification Details 6 second blot, plunged into liquid ethane (GATAN CRYOPLUNGE 3)
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 12000
Reported Resolution (Å) 3.5
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID
Refinement Space
Refinement Protocol
Refinement Target
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector Type KODAK SO-163 FILM
Electron Dose (electrons/Å2) 22
Imaging Experiment
Date of Experiment 2011-01-01
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1100
Maximum Defocus (nm) 2400
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model ?
Nominal Magnification 59000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
RECONSTRUCTION EMAN ?
RECONSTRUCTION FREALIGN ?
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
. each particle