New TopShim run:

version = topshim 2.1.0
date = 2022-01-20
time = 17:27:34.907
command = topshim 

FINE SHIMMING

Parameters:

probe = Z75811_0017
solvent = H2O+D2O_salt
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.7 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 2: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.4 Hz
Iteration 3: FWHM Before= 0.4 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 4: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.5 Hz
Iteration 5: FWHM Before= 0.5 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 6: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.3 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.277

shim changes: 
Z	-6
Z2	-2
Z3	-6
Z4	+18
Z5	+62
Z6	+32
Z7	-370
Z8	-755

duration = 47 sec
completed successfully
finished Thu Jan 20 17:28:20 2022


========================================================================