New TopShim run:

version = topshim 2.1.0
date = 2022-01-19
time = 16:57:11.651
command = topshim 

FINE SHIMMING

Parameters:

probe = Z75811_0017
solvent = H2O+D2O_salt
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.8 Hz
Iteration 2: FWHM Before= 0.8 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 3: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.6 Hz
Iteration 4: FWHM Before= 0.6 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 5: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 6: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.6 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.314

shim changes: 
Z	+11
Z2	+4
Z3	+16
Z4	+85
Z5	-89
Z6	-554
Z7	+677
Z8	+3026

duration = 46 sec
completed successfully
finished Wed Jan 19 16:57:56 2022


========================================================================