New TopShim run:

version = topshim 2.1.0
date = 2022-01-19
time = 17:45:43.403
command = topshim 

FINE SHIMMING

Parameters:

probe = Z75811_0017
solvent = H2O+D2O_salt
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 1.7 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 2: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 1.0 Hz
Iteration 3: FWHM Before= 1.0 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 4: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 5: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 6: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.3 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.326

shim changes: 
Z	-32
Z2	-4
Z3	+2
Z4	-88
Z5	+100
Z6	+699
Z7	-922
Z8	-2855

duration = 46 sec
completed successfully
finished Wed Jan 19 17:46:28 2022


========================================================================