New TopShim run:

version = topshim 2.1.0
date = 2022-01-12
time = 17:54:47.624
command = topshim 

FINE SHIMMING

Parameters:

probe = Z75811_0017
solvent = H2O+D2O_salt
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.4 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 2: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 3: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 4: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 5: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 6: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.25

shim changes: 
Z	+5
Z2	+3
Z3	+11
Z4	-21
Z5	-7
Z6	+173
Z7	+349
Z8	-263

duration = 45 sec
completed successfully
finished Wed Jan 12 17:55:31 2022


========================================================================