New TopShim run:

version = topshim 2.1.0
date = 2021-12-03
time = 23:39:26.322
command = topshim 

FINE SHIMMING

Parameters:

probe = Z75811_0017
solvent = H2O+D2O_salt
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 2: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 3: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 4: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.4 Hz
Iteration 5: FWHM Before= 0.4 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 6: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.2 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.265

shim changes: 
Z	-2
Z2	+6
Z3	+61
Z4	-96
Z5	+55
Z6	-10
Z7	-391
Z8	+344

duration = 47 sec
completed successfully
finished Fri Dec  3 23:40:11 2021


========================================================================